The metering system of the electromagnetic flow meter primary is base on Faraday’s laws of electromagnetic induction, on the channel border which is vertical each other with metering tube axis and line of magnetic field mount one pair detecting electrode, when the conductive liquid move alon metering tube axis conductive liquid cutting the line of magnetic field induce the inductive E.M.F. This EMF detect out by two electrodes on the metering tube, numerical value size is:
E=KBVD
Here: E inductive EMF;
K instrument constant;
B magnetic induction
V average flow velocity in metering tube section;
D metering tube inner diameter;
When measure flow rate fluid flow through magnetic field which is vertical to flow direction, conductive liquid movement induce one EMF in direct proportion to average flow velocity, so that require measured flow liquid have higher than lowest limit conductivity. Its induction voltage signal is detected by two electrodes, and pass through a cable transmit to converter, after through signal treatment and correlative operation, take the integrating flow and the instantaneous delivery indicate on the display screen of the converter. |